| Model | STF1200.50.600 |
| Electrical requirements | 115V, 50/60Hz, 1 phase, 1500 watts |
| Tube | 50 mm OD x 600 mm L, quartz |
| Maximum working temperature | 1200°C (2 hours maximum continuous run time) |
| Constant working temperature | 1100°C |
| Minimum working temperature | Ambient |
| Constant working temperature under vacuum | 1000°C |
| Rated vacuum pressure | 0.001 Pa / 0.0075 millitorr / 0.00001 mbar |
| Rated positive pressure | 0.02 MPa / 150 torr / 3 psi |
| Heating element type | Kanthal (Sweden) A1 2.0 resistance coil wire |
| Thermocouple type | K |
| Refractory lining | Mitsubishi (Japan) 1500-grade fiber alumina |
| Temperature controller | UDIAN 30-segment, single PID |
| Maximum heating rate | < 30°C / min |
| Heating zone length | 8 in |
| Constant temperature zone length | 3 in |
| Temperature controller precision | +/- 1°C |
| Vacuum sealing flange kit | Stainless steel vacuum sealing flanges with one vacuum gauge, two valves, and four thermal ceramic blocks |
| Gas inlet / outlet | 1/4 in hose barb with 1/4 in BSPT |
| Unit dimensions without tube | 25.5 x 17 x 17.75 in (W x D x H) |
| Unit weight | 61 lb without tube/flange/thermal blocks; 68 lb with tube/flange/thermal blocks |
| Shipping weight / dimensions | 120 lb / 31 x 19 x 26 in |
| Compliance | CE |