Please note this unit needs to be plumbed with distilled water or filtered tap water. Water resistance must be in the range of 50K-1M Ohm/cm or conductivity range of 20.0-1.0 uS/cm. If your laboratory cannot support this water quality, purchase the SHEL LAB Water Purification Kit. See the installation and operation manual for more information.
SKU:SHC28 Shelving:6 Supplied (16 Maximum) Relative Humidity Range:Ambient +10% to 95% Jacket Type:Air Temperature Uniformity:+/-0.5°C at 37°C Temperature Range:40°C to 70°C Exterior Dimensions WxDxH in. (cm):42.5 x 37.0 x 85.0 (108.0 x 94.0 x 215.9) Interior Dimensions WxDxH in. (cm):30.2 x 26.0 x 62.0 (76.8 x 66.0 x 157.4) Capacity:28 Cu.Ft. (792.9 L) Maximum Energy Consumption (W):1380/1495
All specifications are determined by using average values on standard equipment at an ambient temperature of (22 ± 3)°C and line voltages within +/-10% of unit type (115V/230V). Temperature specifications follow DIN 12880 methodology. We reserve the right to change specifications at any time.