Please note this unit needs to be plumbed with distilled water or filtered tap water. Water resistance must be in the range of 50K-1M Ohm/cm or conductivity range of 20.0-1.0 uS/cm. If your laboratory cannot support this water quality, purchase the SHEL LAB Water Purification Kit. See the installation and operation manual for more information.
SKU:SHC10 Shelving:3 Supplied (8 Maximum) Relative Humidity Range:Ambient +10% -95% Jacket Type:Air Temperature Uniformity:+/-0.5° at 37°C Temperature Range:40°C to 70° C Exterior Dimensions WxDxH in. (cm):44.3 x 32.8 x 57.0 (112.4 x 83.2 x 144.8) Interior Dimensions WxDxH in. (cm):30.0 x 21.0 x 30.0 (76.2 x 53.3 x 76.2) Capacity:10.0 Cu. Ft. (309L) Maximum Energy Consumption (W):1495
All specifications are determined by using average values on standard equipment at an ambient temperature of (22 ± 3)°C and line voltages within +/-10% of unit type (115V/230V). Temperature specifications follow DIN 12880 methodology. We reserve the right to change specifications at any time.